超快速实时荧光定量PCR仪 HC800
-
Average heating-up rate50℃ —90℃, ≥6.0℃ /s
-
Maximum heating-up rate50℃ —90℃, ≥8.0℃ /s
-
Average cooling rate90℃ —50℃, ≥3.5℃ /s
-
Maximum cooling rate90℃ — 50℃ , ≥ 6.5℃ /s
-
Module control accuracy≤0.5℃
-
Temperature accuracyAbsolute value of difference between measured value and set temperature ≤0.5℃
-
Temperature uniformity±1℃
-
RepeatabilityThe CV of the Ct value shall not be greater than 3%
-
Temperature Accuracy in duration time±5%
-
linearityr≥0.980
-
Fluorescent linearLinear regression coefficient r≥0.990
-
Temperature5℃~40℃
-
Relative humidity20%~80%
-
Atmospheric pressure85.0kPa-106.0kPa
-
Altitude<2000m
-
Use locationIndoor
-
DimensionsL*W*H, mm : 250×190×120
-
Weight3.2kg( With power adaptor 3.8kg)
-
Input voltage100-240VAC
-
Frequency50/60Hz
-
Input power90W
-
Overvoltage levelⅡ
-
Contamination level2
-
Protection level of equipment shellIPX0
-
Sample capacity15-25 μl
-
Fluorescent dyeF1: FAM,SYBR GreenⅠ ;F2: HEX,VIC,JOE,TET,YELLOW;F3: ROX;F4: CY5
-
Chanels4
-
Optical systemLED+MPPC
-
Parameter100-240VAC, 50/60Hz, 1.3-0.6A;15V 6.0A, 90W MAX
-
Grid voltage fluctuation range90-264VAC